Fluorimetric detector model 121M

Fluorimetric detector model 121M
Fluorimetric detector model 121M

The Model 121M fluorimetric detector is designed for the analysis of fluorescent substances, including benz(a)pyrene, aflatoxins, and OPA derivatives of amino acids.

The detector has a unique optical scheme with the use of semiconductor light sources. The quartz cuvette of the cylindrical detector has a volume of 1 µl, which makes it possible to successfully use this device for micro-column HPLC.

Light source monochromatic LED
Fluorescence excitation wavelength 365±5 nm
Range of fluorescence detection wavelengths 400-460 nm
Detectable volume less than 1 µl
Detection limit (for anthracene) < 1·10-14g
Zero signal fluctuation noise level in a cuvette with bidistilled water (at a time constant of 0.5 s) 2, 5x10-5 O. E. F.
Zero signal drift in a cuvette with bidistilled water (with a stable ambient temperature of ± 1°C, with stable operation from 2.5 to 3 hours) 2. 5x10-5 O. E. F./h
Liquid path material PEEK, PTFE, SS316, quartz glass
Inlet and outlet fittings ferrule internal thread 10-32
Time to enter operating mode 10 mins
Power supply, voltage/frequency 220V/50Hz or 110V/60Hz
Power consumption, max 95 VA
Overall dimensions (height, width, depth) 160x330x290 mm
Weight 7.5 kg

O. E. F. - relative unit of fluorescence

Миниатюра схемы