TSC-003 Column thermostat

TSC-003 Column thermostat
TSC-003 Column thermostat

{IMAGE2}

When developing the thermostat for the Steyer-M line, the developers used a number of non-standard design solutions.

The heating element with increased power allows the implementation of heating in a minimum time, and the precise software PID controller provides an optimal heating profile and quick response to changes in external conditions.

A powerful air fan provides fast forced cooling, which allows you to control not only an increase, but also a decrease in the column temperature, and also speeds up the transition to a different temperature control value.

Controlled airflow to all external cavities, including the airflow circuit inside the thermostat cover, allows the thermostat housing to maintain a normal temperature even at maximum column heating, ensuring comfortable work for personnel and surrounding equipment.

Main Features

  • Short heating and cooling times.
  • Stability of maintaining the column temperature is not worse than 0.1 °C at a point and not worse than 0.2 °C over the entire volume of the column.
  • Accuracy of temperature setting is not more than 0.1 °С.
  • Ability to install at least two columns in the thermostat complete with guard column holders.
  • Work in temperature gradient mode.
  • Accelerated response to changes in external conditions and a quick return to operating mode.
  • Convenient and friendly user interface.
  • Rapid temperature drop when the appliance is turned off (forced cooling).
  • Turn the appliance on or off at a specific, predefined time using the built-in timer.
  • Ability to update firmware remotely.
  • Exclusion of the possibility of overheating (temperature rise) above the set one.
  • Availability of software and emergency hardware protection against overheating.
Specifications TS-3.80 TS-3.100*
Temperature Range Troom + 10 °С - 80 °С Troom + 10 °С - 100 °С
Execution type monoblock monoblock
Accuracy of temperature maintenance, °С 0,1 0,1
Temperature setting accuracy, °С 0,1 0,1
Time to reach the mode, minutes, no more 20 30
Cooling time to Troom + 10 °С, minutes, no more 25 35
Heating element power, W 120 120
Type of thermostat cooling forced forced
Number of columns to install 2 2
Number of thermostat steps 5 5
Food 110-240V, 50Hz 110-240V, 50Hz
Programmable turn-on timer there is there is
Remote processor software update there is there is
Type of execution according to CE qualification Industrial Industrial


* TS-3.100 thermostats are available on request